Features
-
Select Programmed or Adjustable
treatment modes.
- Commonly used treatment regimens pre-set for
ease of operation.
- Fully adjustable pulse width, rate, and
amplitude offers flexibility
- Width modulation minimizes patient
accommodation to stimulation
- Dual channels permit treatment of multiple pain
sites
- Advanced engineering increases battery life
- 3 modes of stimulation (normal, burst,
modulation)
- Ergonomically designed for comfortable and
attractive wear.
- American designed and manufactured
- Limited Lifetime Warranty
|
Technical
Specifications
Output
parameters with switch in "adjustable"
setting.
NORMAL
Amplitude: 1-70 milliamps,
adjustable.
Width: 50-450 microseconds,
adjustable.
Pulse Rate: 1-120 Hz,
adjustable.
Cycle Time: N/A.
BURST
Amplitude: 1-70 milliamps,
adjustable.
Width: 50-450 microseconds,
adjustable.
Pulse Rate: 85 Hz, fixed 2 bursts/sec.
Cycle Time: N/A.
PC1
Amplitude: From setting, the
amplitude varies with the pulse width.
Width: From 275 microseconds,
modulates up and down according to the span
setting, 0-100% (max. range of 100-450
microseconds).
Pulse Rate: From 30 Hz, modulates up and
down according to the span setting, 0-100% (max.
range of 5-105 Hz).
Cycle Time: Varies according to
width and rate setting.
PC2
Amplitude: From setting, the
amplitude varies with the pulse width.
Width: From 300 microseconds,
modulates up and down according to the span
setting, 0-100% (max. range of 100-450
microseconds).
Pulse Rate: 1-120 Hz, adjustable (no
rate modulation).
Cycle Time: Varies according to
width and rate.
Output parameters with switch in
"programmed" setting.
NORMAL
Amplitude: 1-70 milliamps,
adjustable.
Width: 110 microseconds.
Pulse Rate: 85 Hz, fixed.
Cycle Time: N/A.
BURST
Amplitude: 1-70 milliamps,
adjustable.
Width: 100 microseconds.
Pulse Rate: 85 Hz, fixed 2 bursts/sec.
Cycle Time: N/A
PC1
Amplitude: From setting, the
amplitude varies with the pulse width.
Width: Modulates from 400
microseconds to 150 microseconds.
Pulse Rate: Modulates from 5-105 Hz.
Cycle Time: 12 sec. total time.
PC2
Amplitude: From setting the
amplitude varies with the pulse width.
Width: Modulates from 450
microseconds to 150 microseconds.
Pulse Rate: 65 H, fixed (no rate
modulation)..
Cycle Time: 10 sec. total time.
|